Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)

Product Details
Customization: Available
Type: Defibrillator
Certification: CE, ISO13485
Manufacturer/Factory, Trading Company
Diamond Member Since 2009

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Importers and Exporters
The supplier has import and export rights
High Repeat Buyers Choice
More than 50% of buyers repeatedly choose the supplier
Years of Export Experience
The export experience of the supplier is more than 10 years
Experienced Team
The supplier has 6 foreign trading staff(s) and 4 staff(s) with over 6 years of overseas trading experience
to see all verified strength labels (9)
  • Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)
  • Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)
  • Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)
  • Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)
  • Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)
Find Similar Products
  • Overview
  • Product Description
  • Packaging & Shipping
  • Company Profile
Overview

Basic Info.

Model NO.
THR-DM-900A
Group
Adult
LCD Display
With LCD Display
Display
8 Inch TFT Color Display
Energy Selection
2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 Joule
Charge Time
Less Than 7 Seconds
Charging Prompts
Voice and Visual Prompts
Output Waveform
Biphasic Technology
Standard Parameter Is ECG
I II III AVR Avl Avf V
MOQ
1
Printer
Built-in Printer
Transport Package
Standard Export Packing
Specification
CE
Trademark
THR
Origin
China
Production Capacity
50000piecs/Year

Product Description

Product Description

Defibrillator with Multi parameter Patient Monitor
Model:THR-DM-900A
Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)

 

with voice alarm that meets your clinical needs from Basic Life Support (BLS)

to Advanced Cardiac Life Support (ACLS).
Manual defibrillator 2J-360J ,AED 50-200J
Multi-parameter patient monitor ECG.SpO2.NIBP.EtCO2.Pacer.


Available colour
Grey . Blue . White

Defibrillators benefit patients who experience a sudden heartbeat (arrhythmia). In these cases
Defibrillators are an important tool for anyone who is near a victim suffering from a ventricular arrhythmia.
The arrhythmia causes the patient to lose consciousness. The expected outcome of such an arrhythmia is either cardiac arrest or stroke.
Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)


Features of THR-DM-900A
High resolution 8 inch TFT color display
Energy selection: 2,5,7,10,20,30,50,70,100,150,200,300,360 Joule
Defibrillator device voice alarm
Charge time: Less than 7 seconds with a new, fully charged battery
Energy display: monitor display shows both selected and delivered energy
Charging prompts: Voice and visual prompts
Electrode impedance measurement range: 0-250 ohms
Manual mode
Built-in printer
Reusable standard external adult and pediatric paddles (pediatric paddles integrated)
AED function: Automatic analysis and chargingx3 with programmable automatic energy level selection
Output waveform: Biphasic technology
Standard Parameter is ECG: I II III AVR AVL AVF V


Specification of THR-DM-900A
Defibrillator
Waveform Details:

The tables below detail the waveforms that the Defi® Xpress delivers when connected to resistive loads of 25, 50, and 100 ohms and set to its maximum output power.
The waveforms are identified by the peak current (Ip) values DEFI XPRESS for two-phase operation: Duration of the first output phase (tphase1) and Duration of the second output phase (tphase2).
The specified values are within 10%.
Waveform: Truncated exponential curve Biphasic

Output energy accuracy:
+/- 10% or 1J (whichever is greater) at 50 ohms
+/- 15% or 1J (whichever is greater) at 25 to 100 ohms

Energy Selection:
External: 2 5 7 10 20 30 50 70 100 150 200 300 360 J.

Recharge time:
< 06 seconds at 200 J
 < 12 seconds @ 360 J
< 15 seconds @ 360 J 

Charge level indication: Audible and graphic.

Output :
Adult paddle, pediatric adapter and multipurpose hands-free adapter available.
Synchronizer : Delivers power within 60 msec.
Disarm: Front panel switch.

Monitor/Display
Input:
3-wire 5-core patient cable
Size: 8 inch
Type: Color (TFT)
Display resolution: 800 x 600 pixels.
Passing speed: 12.5mm/s25mm/s50mm/s
Lead selection: I II III AVR AVL AVF V.
Frequency response: (user selectable).
0.7 to 20 Hz Operating mode
0.3 to 40 Hz Monitoring mode
0.05 to 150 Hz Analysis mode
0.5 to 30 Hz AED mode
Common mode rejection: > =60dB
Heart rate monitor: 20 to 300 BPM.
Heart rate alarms: user selectable.
Size: 0.125 0.25 0.5 1 2 4 cm/mv and auto ranging.
ECG output: 1 V/mv.
Heart Rate Response to Irregular Rhythm: (AAMI EC13-1992 Section 4.1.2.1.e.)

ECG Monitoring
Patient Connection: 3-lead ECG cable, or 5-lead cable, paddles.
Lead Selection: Display on monitor, paddles, I, II, III, AVR, AVL, AVF, V,
ECG Size: 0.25, 0.5, 1, 1.5, 2, 4cm/mV display on monitor
Heart Rate: 20-300 BPM
Heart Rate Alarm, On/Off displayed on monitor, user-selectable.
Smart Alarms: Beeper/Voice prompts indicate shock able rhythm.

Alarms
Heart Rate Spo2 Alarm:
Audible: 5-pulse 800 tone with a PW of 150 msec, a PRI of 225 msec, and a repeat interval of 10 seconds.
Visual: The heart rate alarm flashes the displayed heart rate at 2 Hz. This indicator is located at the top of the display and is 0.4" high and 0.28 to 0.840" wide, depending on the number of digits in the heart rate. It is amber, black, white, or the color of the display text.

Line fault alarm:
Audible: triplet tone with 3 pulses and 500 Hz with a PW of 200 msec and a PRI of 310 msec. If the HR alarm is set or the pacemaker is turned on, the line fault tone repeats at a 20 second repetition interval.
Visual: When a line fault occurs, the message "LEAD FAULT" is displayed on the trace along with a dashed line the width of the trace. The text is 0.18" high and is the color of the display text. The dashed line is 5" long and has the width of the display. The signal is not modulated.

NIBP (options):
Method: automatic oscillation
Working mode: manual/automatic
Measurement: adult: 10-250mmHg
Ement: pediatric: 10-200mmHg
Ement: neonatal: 10-135mmHg

End-tidal CO2 (EtCO2):
Safety EtCO2:BF type
Current consumption 12.0V±0.2V DC
Current consumption 100mA (typical with pump on) 200mA (extreme condition)
Operating temperature range 5~50ºC
Shipping/storage temperature range -20~+70ºC
Relative humidity 0-85% (non-condensing)
Dimensions 96×91×36mm
CO2 range 0 to 13%(0-98.8mmHg@BTPS)
CO2 accuracy < 5.0% CO2 (at BTPS) ± 3 mmHg > 5.0% CO2 (at BTPS) < 10% of reading
Respiratory rate 3~60 BPM
Measurement accuracy ±1 BPM
Warm-up time 10 seconds
Detector response time: 10mS,
System: depends on implementation, pneumatics and water separation technology
Flow control 50~250ml/min
Interface Highly configurable serial digital interface (TTL/RS232)
Automatic offset calibration time and temperature dependent and as required

Pacemaker (options)
Type: VVI
Pulse type: rectangular, constant current
Pulse amplitude: 0 to 180 mA ± 10% or 5 mA (whichever is greater).
Pacing frequency: Variable from 30 ppm to 180 ppm ±1.5% (increments or decrements by a value of 2 ppm)
Multi-Functional Electrode: (MFE) Pads: Multipurpose defibrillation/pacing electrodes
Pause: pacing pulse rate is reduced by a factor of 4 upon activation
Refractory period: NC (VVI requirements do not have this specification)
Pulse width: (20+ 1.5 ms)
Output protection: Fully defibrillator protected and isolated

Recorder
Type: High resolution thermal array.

Recording: time date ECG lead ECG gain heart rate defibrillation and treatment summary ACLS events.
 
Standard:
ECG Cable (5 Leads) 1 pcs
AED Pad Adapter  1 pcs
AED Pad 1 pcs
Ground Cable 1 pcs
Power Cable 1 pcs
Thermal Paper Role 1 pcs
Cleaning Cloth 1 pcs

Options:
AED Pad for Child
ECG Cable (3 Leads)
ECG Alligator Clips (Metal)
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor) 
SpO2 Cable (Neonatal soft sensor) 
SpO2 Cable (Pediatric hard shell sensor) 
SpO2 Cable (Pediatric Soft shell sensor)
SpO2 Cable (Child finger clip)
SpO2 Cable (multi-site Y type)
SpO2 External Cable
NIBP Cuff (infant 8 - 14 cm 5/Bag)
NIBP Cuff (Child 13 - 20 cm 5/Bag)
NIBP Cuff (Small Adult 18 - 26 cm 5/Bag)
NIBP Cuff (Adult 26 - 35 cm 5/Bag)
NIBP Cuff (Adult Long 29 - 38 cm 5/Bag)
NIBP Cuff (Large Adult 32 - 42 cm 5/Bag)
NIBP Cuff (Large Adult Long 35 - 44 cm 5/Bag)
NIBP Cuff (Thigh 42 - 50 cm)
NIBP Tube 150cm
Paper Role
Defibrillator Wall mounts (3 Models)
Defibrillator Trolley (5 Models)
Defibrillator Case
EtCO2 Module
 
Please feel free to contact us for more discount price!
Packaging & Shipping

 

Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)
Company Profile

Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)Hospital First-Aid Portable Automated External Biphasic Defibrilator (THR-DM-900A)

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier
Product Groups
More