Medical Device Regulatory Type: | Type 3 |
---|---|
Transport Package: | Standard Export Carton or Wooden |
Specification: | AC220V-230V/50HZ |
Trademark: | THR |
Origin: | Jiangsu, China (Mainland) |
Customization: |
---|
Suppliers with verified business licenses
Brief description of incubator neonatal:
1) Two control modes: air model and baby mode, controlled by micro-computer;
2) Set Temperature, Air temperature, Baby temperature are displayed separately;
3) Natural air flow humidity;
4) Inclination of bassinet is adjustable;
5) Failure alarm indication;
6) RS-232 connector;
7) THR-II90AB incubator neonatal is configured with fixed Neonate Bilirubin Phototherapy Equipment (with blue fluorescent lamps)
Standard Configuration of incubator neonatal:
The hood, the base, the mobile cabinet, the controller, I.V pole, Shelf.
With Neonate bilirubin phototherapy equipment(fixed model)
Specification of incubator neonatal:
Power supply |
AC220V-230V/50HZ |
Max power output |
850VA |
Mode of control |
Air mode and baby mode controlled by micro-computer |
Air temperature control range |
25°C-37°C (override model: 37-39°C) |
Baby temperature control range |
34-37 °C (override model: 37-38°C) |
Accuracy of skin temperature sensor |
≤ 0.3°C |
Temperature variability |
≤ 0.5°C |
Temperature uniformity |
≤ 0.8°C (mattress is on the horizontal position) |
Inclination of the bassinet |
± 12° |
Noise inside hood |
≤55dB(A) (environment noise ≤ 45dB(A)) |
Failure alarm |
Over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on. |
Environmental Temperature:
Operating range: 20°C -30°C
(set temperature should be higher 3°C than the environment temperature)
Ambient air movement rate: ≤ 0.3m/s
Suppliers with verified business licenses