The sterilizer uses microprocessor intelligent control, automatic door locking, human-machine interface, easy to operate, safe and reliable.
Product name | Autoclave Sterilization Hospital Steam Disinfectant Machine |
Sterilization roon size(cm) |
Φ31.9*61.5 |
Outline dimension(cm) | 94*59*57.5 |
Product form |
Desktop |
Heating mode |
electrical heating |
Power supply |
AC220V,50Hz |
Sterilizing chamber material |
The sterilizing chamber is 304 anticorrosive stainless steel with thickness ≥ 3mm, safety door thickness ≥ 15mm, hydraulic test ≥ 0.8mpa. which exceeds the national standard of YY0646-2008 design. |
Outer box material |
Carbon steel (1.5mm thick) baking powder |
Work environment |
0-50ºC |
Working pressure |
-80kpa~200kpa |
Sterilization humidity | ≥80% |
Maximum pressure resistance | -120kpa~260kpa |
Sterilization temperature | 121ºC-135ºC adjustable |
Vacuum mode | Three pre-vacuum to ensure faster heating, sterilization more thorough. |
Over voltage protection program | When the sterilizing chamber pressure reaches 220 ±5kPakPa, there are cues and automatic pressure relief. |
Pressure sensor | The inner pressure sensor automatically detects the pressure inside the cavity to ensure the accuracy and efficiency of the pressure. |
Temperature sensor | The built-in temperature sensor automatically detects the temperature inside the cavity to ensure that the temperature in the cavity is within the accurate range of ±3 ºC. |
Sterilization mode |
Bare equipment, cotton yarn, wrapping equipment, plastic and rubber equipment products, the default working temperature is 135 ºC. |
Sterilization time at 135 ºC |
36 minutes(Preheat 12 minutes, vacuum 4 minutes, sterilization 4 minutes, dry 16 minutes) |
Sterilization time at 122 ºC |
56 minutes(Preheat 8 minutes, vacuum 4 minutes, sterilize 5-20 minutes, dry 20 minutes) |
Alarm system |
High and low temperature alarm system |
Category |
Class II medical devices |