| Description |
| FHR(Fetal Heart Rate) |
|
Method: Ultrasound Doppler |
|
Measuring Range: 60-200bpm |
|
Precision: ≤±2bpm |
|
Ultrasound Frequency: 2.5 MHz |
|
Ultrasound output: 5mW/cm2 |
|
Ultrasound Task Measurement:Impulse Wave |
|
Alarm Indication range: upper limit:66-200bpm |
|
Lower limit: 65-199bpm |
|
TOCO (Uterine Contraction Pressure) |
|
Method: Outside Measurement |
|
Measuring Range: 0-100 unit |
|
Non-linearity Precision: ≤±10% |
| Quickening Sign |
| Quickening Sign: Artificial and Automatic |
|
SPO2 |
|
Sensor:Two-waveform LED(light emitting diode) |
|
SPO2 measuring range:0-100% |
|
Resolution:1% |
|
Precision: ≤±2%(90%-100%);≤±4%(70%-89%);≤±6%(40%-69%) |
|
Alarm Indication range:Upper limit 1-100%;Lower limit 0-99% |
|
Range of pulse rate:20-250 bpm |
|
Pulse alarm indication range:Upper limit 22-250 bpm;Lower limit 20-248 bpm |
|
NIBP |
|
Method: Intelligent Oscillation |
|
Cuff inflating time: 30s (standard adult cuff 0-250 mmHg) |
|
Deflating time: average90sec |
|
Deflating time when measurement stop:2sec(standard adult cuff 250-150mmmHg) |
|
Initial inflating pressure:180mmHg(adult); 120mmHg(infant) |
|
Maximum allowed inflating pressure:300mmHg(40Kpa,over this value, safety Value deflating automatically) |
|
Measuring Mode:Manual measurement, adult mode |
|
Measurement Intermission:0-240min adjustable |
|
Measurement parameters:Systolic pressure,Diastolic pressure,Mean pressure |
|
Alarm indication range:Upper limit: 22-250mmHg;Lower limit: 20-248mmHg |
|
Alarm Indication |
|
Alarm indication range of FHR, UCP.HR, NIBP SPO2 are settable. |
|
Sound and light alarm are available. |
|
|
| Other Specification |
|
Power supply:220V±22V,50Hz±1Hz |
|
Power consumption: ≤110VA |
|
Display mode: 12.1 inch TFT LCD display |
|
Alarm mode: Sound and Light |
|
Working condition: Ambient temperature: 5-40 |
|
Relative humidity: ≤80% |
|
Atmospheric pressure: 86-106Kpa |
|
Storing condition: Ambient temperature: -10-40 |
|
Relative humidity: ≤80% |
|
Keep away from acrid gas |
|
Good draught |
|