Features:
1. Two temperature control modes: Air mode and Baby mode controlled by micro-computer
2. Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance
3. Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
4. Failure alarm indication
5. Can be equipped with > 37°C temperature set function
6. Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it
7. Natural air flow humidity
8. The height of whole unit can be adjusted
9. Oxygen supply system
10. Observation lamp
Specification:
AC power requirement |
AC220V- -230V/50Hz or AC110-120V/50-60Hz, 400VA |
DC Power requirement |
DC12V/10A or DC24V/6A |
Control Mode |
Air mode and Baby mode controlled by micro- computer |
Air temperature control range |
25°C - 37°C(Override mode 37°C-38°C) |
Baby temperature control range |
34°C- -37°C(Override mode 37-37.5°C) |
Skin temperature sensor accuracy |
±0.2°C |
The temperature fluctuation range between incubator temperature and average incubator temperature (under stable temperature condition) |
≤1.0°C |
Temperature uniformity |
≤1.5°C(mattress is on the horizontal position) |
Noise inside hood |
≤52 dB(A) |
Failure alarm |
over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and SO on. |
Battery continuous working time |
90min (Set temperature is 36°C- Ambient temperature is 15°C) |
Product Packaging:
Package: Main body and Cabinet are packed into two cartons
Dimension (Main body): 1090mm * 655mm * 760mm
(Cabinet): 1380mm * 610mm * 400mm
Gross Weight (Main body): 80Kg
(Cabinet): 53Kg