Customization: | Available |
---|---|
Type: | Infusion Pump |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Item Name | Infusion Pump |
Model | THR-IP900 |
Screen Size | 4.3 inch |
Standard Parameters | Remote Control, Drug library, History Records, Heating function |
Working Mode | Drop Mode Volume Mode Time Mode |
Customized LOGO | Support |
Certificates | ISO & CE |
Guarantee | 12 months |
Dimensions | 23*24*33.5cm |
Net Weight | 2.3kg |
Gross Weight | 3.1kg |
Brochure Download | PDF File |
KVO Rate:
|
1 ml/h (1~300 ml/h) 2 ml/h (100~300ml/h) 3 ml/h (greater than 300 ml/h) |
Adjustable Drop Rate Range
|
1~400 drops/min (step: 1 drop/min)
|
Adjustable Volume Rate Range
|
1~1200 ml/h (when 1~99.9 ml/h, step: 0.1 ml/h; greater than 99.9 ml/h, step: 1 ml/h)
|
Occlusion Alarm Threshold
|
High: 800 mmHg±200 mmHg (106.7kPa±26.7kPa) Medium: 500 mmHg±100 mmHg (66.7kPa±13.3kPa) Low: 300 mmHg±100 mmHg (40.7kPa±13.3kPa) |
Mechanical Precision
|
within ±2%
|
Purge Rate:
|
800 ml/h
|
Mechanism
|
Peristaltic pump
|
Infusion Volume Range
|
1 ml~9999.9 ml (step: 0.1 ml)
|
Maximum Accumulated Volume
|
9999.9 ml (step: 0.1 ml)
|
Flow Rate Accuracy
|
Within ±3% (using calibrated high quality IV set)
|
Time Range:
|
1 min~9999 min (step: 1 min)
|
Alarm
|
Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.
|
Air Bubble Detector
|
Method: ultrasonic wave, sensitivity: upon 25 μL
|